24.2.1Medical physics and astronomy (A Level)

Production and Use of X-rays

5 min read 11 parts Lesson 8 of 22 1 quiz

0%read
Lesson summary

24.2.1 Production and Use of X-rays — AS PHYSICS 9702 notes: Describe how X-rays are produced, how the tube controls and spectrum work, and how images are improved while limiting dose.X-rays — They are the short-wavelength, high-frequency part of the electromagnetic spectrum, wit…

  • Properties
  • Accelerating p.d
  • Bombardment
  • Outer electrons moving down
  • Maximum energy
  • Minimum wavelength

Section 1 of 11X-rays

X-rays
  • They are the short-wavelength, high-frequency part of the electromagnetic spectrum, with wavelengths from about 10−8 m to 10−13 m.
Production
  • They are produced when fast-moving electrons decelerate rapidly and transfer their kinetic energy to photons.
Properties
  • X-rays travel in straight lines at the speed of light, are unaffected by electric and magnetic fields, blacken film, cause fluorescence and ionise matter.
Photons
  • X-rays are uncharged photons, not charged particles.

Section 2 of 11The X-ray tube

Cathode
  • A heated filament releases electrons by thermionic emission.
Accelerating p.d.
  • A large p.d. of typically 30 kV to 150 kV accelerates the electrons towards the anode.
Vacuum
  • The tube is evacuated so electrons do not collide with air molecules.
Target
  • A metal anode, commonly tungsten, has a high melting point and high atomic number.
Heat
  • Less than 1% of the electron energy becomes X-rays and the rest becomes heat.
  • High-power tubes use a rotating anode, cooled by oil or water, to spread the heat.

Section 3 of 11Production mechanism

1Bombardment
  • Electrons striking the target lose kinetic energy by transferring it to photons.
2Outer electrons moving down
  • Outer-shell electrons in the target atoms drop into vacancies in lower levels and release X-ray photons.
3Maximum energy
  • An electron accelerated through V gains Emax = eV, the maximum energy of a photon.
4Minimum wavelength
  • Emax = eV = hfmax = hc/λmin.
  • fmax = eV/h and λmin = hc/(eV).

Section 4 of 11Controlling the beam

Tube current
  • It is set by the filament temperature, and raising it increases the number of electrons hitting the target per second.
  • The intensity (photons per second) rises without changing photon energy.
Tube voltage
  • It sets the maximum photon energy and hence the penetrating power.
  • Higher V gives harder X-rays with greater penetration.
  • Lower-energy photons are called soft X-rays.
Hardness and intensity
  • Hardness is photon energy and intensity is photons per second, so do not confuse them.

Section 5 of 11The X-ray spectrum

The X-ray spectrum
Continuous spectrum
  • Also called bremsstrahlung, it has all photon energies up to eV.
  • Electrons decelerate in different ways, some in one event and others in many small steps.
Sharp cut-off
  • At λmin, since each electron produces one photon and gives up all its energy in one collision at most.
Characteristic lines
  • An incident electron removes an inner-shell electron and an outer electron falls into the vacancy.
  • The photon energy equals the difference between the shell levels.
  • The lines depend only on the target material and not on the tube voltage, if V is high enough to excite them.
Cut-off wavelength
  • It depends on V and not on the target material.
Effect of current
  • Raising it increases the spectrum height without moving λmin or the lines.

Section 6 of 11Improving the image

Contrast
  • The difference in the degree of blackening between structures.
  • Use the correct hardness, with hard X-rays for bones and soft X-rays for tissue.
  • Use a contrast medium.
Sharpness
  • How well-defined the edges of structures are.
  • Use a narrower X-ray beam.
  • Reduce scattering with a collimator or lead grid.
  • Use a smaller pixel size.

Section 7 of 11Reducing exposure and safety

Hazard
  • X-rays are ionising and can damage cells and DNA, potentially causing cancer.
Aim
  • Give the lowest dose that produces an adequate image.
Aluminium filter
  • It absorbs long-wavelength X-rays that would be absorbed by the body without helping the image.
Collimator and shielding
  • Lead collimators limit the area exposed and lead shields protect other parts of the body.
Staff
  • Staff stand behind lead screens and wear dosimeters to monitor exposure.
Risk and benefit
  • The benefit must outweigh the small dose.

Section 8 of 11Other uses

Imaging
  • Dense materials of high proton number, such as bone, absorb more than soft tissue.
Other applications
  • Security scanning, inspection of welds and X-ray diffraction of crystals.
Radiotherapy
  • High-energy X-rays are directed at tumours to kill cancer cells.

Section 9 of 11More from the specification

More from the specification
Exponential attenuation of X-ray beams
  • As an X-ray beam passes through matter, its intensity decreases exponentially according to the linear attenuation model equation I = I0 e-mu * x,
  • where mu is the linear absorption (attenuation) coefficient and x is tissue thickness.
Tissue contrast and linear attenuation coefficients
  • Different biological tissues have distinct linear absorption coefficients,
    • (e.g., dense bone has a much higher mu ~ 12 cm-1 compared to soft muscle tissue mu ~ 0.20 cm-1
  • producing high radiographic contrast on film or digital detectors.
Computed tomography (CT scans) vs plain radiographs
  • A single plain X-ray radiograph produces a 2D projection with overlapping structures,
  • whereas computed tomography takes multiple X-ray projections around the patient to reconstruct 3D slice images.
Radiation dosage and patient safety
  • Because CT scans combine multiple beam angles to generate sectional image slices, patient exposure to ionizing radiation is significantly higher than from a single standard planar X-ray examination.
Exponential attenuation of X-ray beams
  • As an X-ray beam passes through matter, its intensity decreases exponentially according to the linear attenuation model equation I = I0 e-mu * x,
  • where mu is the linear absorption (attenuation) coefficient and x is tissue thickness.
Tissue contrast and linear attenuation coefficients
  • Different biological tissues have distinct linear absorption coefficients,
    • (e.g., dense bone has a much higher mu ~ 12 cm-1 compared to soft muscle tissue mu ~ 0.20 cm-1
  • producing high radiographic contrast on film or digital detectors.
Computed tomography (CT scans) vs plain radiographs
  • A single plain X-ray radiograph produces a 2D projection with overlapping structures,
  • whereas computed tomography takes multiple X-ray projections around the patient to reconstruct 3D slice images.
Radiation dosage and patient safety
  • Because CT scans combine multiple beam angles to generate sectional image slices, patient exposure to ionizing radiation is significantly higher than from a single standard planar X-ray examination.
Computed tomography
  • An imaging technique that combines X-ray images of slices taken at many angles into a three-dimensional image.
Attenuation
  • The reduction in intensity of radiation as it passes through matter.
Radiation
  • Transfer by electromagnetic waves, requiring no medium
Absorption
  • The conversion of wave energy to internal energy of the medium.
Linear
  • A graph that is a straight line through the origin, so resistance is constant.
Tomography
  • Imaging in slices.
Distance
  • Total length of path travelled, a scalar.

Section 10 of 11Key terms

TermMeaning
X-rayA high-frequency electromagnetic wave produced when fast electrons are stopped by a metal target.
Thermionic emissionThe release of electrons from a heated metal surface.
BremsstrahlungThe continuous X-ray spectrum produced when electrons decelerate in the target.
Characteristic linesSharp X-ray lines at wavelengths determined only by the target material.
ContrastThe difference in the degree of blackening between structures in an image.
SharpnessHow well-defined the edges of structures are in an image.

Section 11 of 11Exam technique

1
Saying a higher tube voltage increases the number of electrons — it increases their energy; filament current controls the number.
2
Stating characteristic lines move when the voltage changes — they depend on the target element only.
3
Forgetting that most electron energy becomes heat — less than 1% becomes X-rays.
4
Confusing hardness with intensity — hardness is photon energy and intensity is photons per second.
5
Describing X-rays as charged particles — they are uncharged photons.