Detecting Gamma-Rays from PET Scanning

5 min read 11 parts Lesson 13 of 22 1 quiz

0%read
Lesson summary

24.3.3 Detecting Gamma-Rays from PET Scanning — AS PHYSICS 9702 notes: Explain how scintillators, photomultipliers and coincidence detection locate tracer in a PET scan and form an image.Layout — The patient lies stationary in a tube surrounded by a ring of detectors.Images — Ima…

  • Absorption
  • Faint light
  • Photocathode
  • Pulse size
  • Different jobs
  • Time window

Section 1 of 11The detector ring

The detector ring
Layout
  • The patient lies stationary in a tube surrounded by a ring of detectors.
Images
  • Images of slices of the body show the position of the radioactive tracer.
Two parts
  • Each detector has a crystal scintillator and a photomultiplier.

Section 2 of 11Scintillator

Absorption
  • A gamma photon hitting the crystal excites an electron to a higher energy state.
Cascade
  • The excited electron excites more electrons as it travels through the crystal.
Light
  • When the excited electrons fall back, the lost energy is emitted as visible light photons.
Materials
  • Examples are bismuth germanate and lutetium oxyorthosilicate.

Section 3 of 11Photomultiplier

Faint light
  • The scintillator's photons are very faint and must be amplified.
Photocathode
  • The light releases electrons from a photocathode.
Dynodes
  • A series of dynodes multiplies the electrons in a cascade, giving a measurable electrical signal.
Pulse size
  • It shows the photon energy, so electronics accept only pulses near 511 keV and reject scattered photons.
Different jobs
  • The scintillator converts gamma to light and the photomultiplier converts light to an electrical signal.

Section 4 of 11Coincidence detection

Time window
  • Electronics look for two pulses from different detectors within a few nanoseconds.
Accepted event
  • Such a coincidence is taken as one annihilation.
Single pulses
  • A pulse with no partner is ignored, since it may be background or a photon whose partner was absorbed.
Line of response
  • Each coincidence defines a straight line joining the two detectors, along which the annihilation occurred.
  • The photons travel in a line because of conservation of momentum.
No collimator
  • Unlike a gamma camera, PET needs no lead collimator, which gives it high sensitivity.

Section 5 of 11Locating the annihilation

Intersections
  • The computer finds where lines of response intersect, the regions with the highest activity.
Time of flight
  • If the annihilation is exactly mid-way, both photons arrive together.
  • If it is a distance x closer to one detector, the path difference is 2x, so Δt = 2x/c.
  • The displacement from the centre is half of cΔt.
Accuracy
  • Timing to a few hundred picoseconds locates the event to within several centimetres along the line.

Section 6 of 11Image construction

Counts
  • A scan collects millions of coincidence events over several minutes.
Voxels
  • The computer divides the body into voxels and finds the tracer concentration from lines passing through each.
Brightness
  • The more photon pairs formed at a point, the more tracer is present, appearing as a brighter point.
False colour
  • Hot areas show high uptake, such as a tumour, and cold regions show low activity.
Attenuation correction
  • Data are corrected for absorption of gamma rays, often using a CT scan in the same session.

Section 7 of 11Strengths and limitations

Strengths
  • It shows metabolic function that other imaging cannot, can detect disease early and is quantitative.
Limitations
  • It is costly, needs a nearby cyclotron, gives a dose from the tracer and any CT, and has a resolution of several millimetres.
  • Resolution is limited by positron range and detector size.
Patient
  • The patient must stay still.
Activity
  • Enough counts are needed for an image, but the activity should be as low as possible.

Section 8 of 11Comparison with other techniques

Comparison with other techniques
Function
  • PET shows how an organ is working rather than how it is built.
Radiation
  • It uses ionising radiation, unlike ultrasound.
Resolution
  • It costs more and has lower spatial resolution than CT.

Section 9 of 11More from the specification

More from the specification
Positron Emission and Radiotracer Decay
  • In Positron Emission Tomography (PET), a patient is injected with a biologically active radiotracer incorporating an isotope with a short half-life that undergoes beta-plus (β⁺) nuclear decay, emitting a positron inside body tissues and metabolic organs.
    • (such as fluorodeoxyglucose, ¹⁸F-FDG)
Positron-Electron Annihilation and Gamma Photon Emission
  • When the emitted positron encounters an electron in surrounding tissue, they undergo mutual annihilation;
    • their rest masses convert into electromagnetic energy, releasing two identical gamma-ray photons travelling in opposite directions (180° apart) governed by the mass-energy equation E = m·c² = h·f.
Coincidence Detection and PET Image Reconstruction
  • Ring detectors surrounding the patient detect pairs of gamma photons arriving simultaneously;
    • calculating the slight difference in arrival times (time-of-flight) pinpoints the exact site of annihilation,
  • allowing computer algorithms to map metabolic activity and disease across patient tissues.
Positron Emission and Radiotracer Decay
  • In Positron Emission Tomography (PET), a patient is injected with a biologically active radiotracer incorporating an isotope with a short half-life that undergoes beta-plus (β⁺) nuclear decay, emitting a positron inside body tissues and metabolic organs.
    • (such as fluorodeoxyglucose, ¹⁸F-FDG)
Positron-Electron Annihilation and Gamma Photon Emission
  • When the emitted positron encounters an electron in surrounding tissue, they undergo mutual annihilation;
    • their rest masses convert into electromagnetic energy, releasing two identical gamma-ray photons travelling in opposite directions (180° apart) governed by the mass-energy equation E = m·c² = h·f.
Coincidence Detection and PET Image Reconstruction
  • Ring detectors surrounding the patient detect pairs of gamma photons arriving simultaneously;
    • calculating the slight difference in arrival times (time-of-flight) pinpoints the exact site of annihilation,
  • allowing computer algorithms to map metabolic activity and disease across patient tissues.
Nuclear equation
  • Equation in which nucleon number and charge are conserved.
Half-life
  • Time for half the nuclei of an isotope to decay.
Frequency
  • Revolutions per second, f = 1/T.
    What the symbols mean: T = period, the time for one complete wave (s) · f = frequency (Hz)
E
E hf = mec2
What the symbols mean: E = energy of one photon (J) · h = Planck constant (6.63 × 10−34 J s) · f = frequency (Hz)
E = hf
What the symbols mean: E = energy of one photon (J) · h = Planck constant (6.63 × 10−34 J s) · f = frequency (Hz)

Section 10 of 11Key terms

TermMeaning
ScintillatorA crystal that emits visible light when a gamma-ray photon is absorbed.
PhotomultiplierA device that amplifies faint light into an electrical signal using a photocathode and dynodes.
Coincidence detectionRegistering two pulses from different detectors within a few nanoseconds as one annihilation.
Line of responseThe straight line joining two detectors that register a coincidence, along which the annihilation occurred.
Time of flightUsing the difference in arrival times of two photons to find the position of the annihilation along the line.

Section 11 of 11Exam technique

1
Thinking PET uses a lead collimator in front of the detectors — coincidence timing replaces it.
2
Saying one detector firing is enough to register an event — two simultaneous detections are needed.
3
Using the full path difference as the displacement — the displacement is half of cΔt.
4
Confusing a scintillator with a photomultiplier — the scintillator turns gamma into light and the photomultiplier turns light into an electrical signal.
5
Forgetting that the scanner shows function rather than anatomy — PET gives metabolic information.