1.3.1Development of Practical Skills in Physics
Using Appropriate Instruments and Techniques
5 min read 8 parts Lesson 18 of 29 1 quiz Virtual lab
1.3.1 Using Appropriate Instruments and Techniques — OCR A Level Physics A notes: Choose a suitable measuring instrument for a task, state its resolution and uncertainty, and use good technique to reduce systematic and random errors.Every measurement starts with a choice of instrument.
- systematic error
- zero error
- repeat readings
- Avoid parallax
- Check the zero
- Measure large multiples of a small quantity
Section 1 of 8Choosing an instrument
Every measurement starts with a choice of instrument.
The best instrument is the one whose resolution is small compared with the quantity being measured.
Other practical factors also matterthe instrument must not disturb what is being measured (a large thermometer in a small volume of water draws heat from it), and it must respond quickly enough for the change being studied.
Section 2 of 8Accuracy, precision, resolution and uncertainty
These words have precise meanings in practical physics and examiners expect them to be used correctly.
For an analogue scale the reading uncertainty is usually taken as half the smallest division, or the full smallest division if the scale is hard to read.
Section 3 of 8Systematic and random errors
A systematic error shifts every reading by the same amount or the same factor in one direction.
A random error causes readings to scatter either side of the true value, for example human reaction time when using a stopwatch.
Section 4 of 8Good technique
Section 5 of 8Percentage uncertainty
The percentage uncertainty compares uncertainty with the size of the measurement:
percentage uncertainty = (absolute uncertainty / measured value) × 100%
It is the percentage uncertainty that tells you whether a measurement is good enough.
This lesson links directly to analogue reading (1.3.2), digital instruments (1.3.3) and the calipers and micrometer (1.3.5), where the details of resolution and zero error for each tool are developed.
Section 6 of 8Worked examples
A student measures the diameter of a wire with a micrometer and gets 0.62 mm. With the jaws closed it reads 0.03 mm. The resolution is 0.01 mm. Give the corrected diameter and its percentage uncertainty.
1Zero error is +0.03 mm, so corrected diameter = 0.62 − 0.03 = 0.59 mm.2Take the uncertainty as ± 0.01 mm.3Percentage uncertainty = 0.01 / 0.59 × 100% = 1.7%.4Answer: (0.59 ± 0.01) mm, 1.7%.The thickness of a stack of 50 identical card sheets is 6.5 mm measured with a ruler of resolution 1 mm (take the uncertainty as ± 1 mm). Find the thickness of one sheet with its uncertainty.
Worked examplesOne sheet: 6.5 / 50 = 0.13 mm.Uncertainty in one sheet = 1 / 50 = 0.02 mm.Answer: (0.13 ± 0.02) mm. The percentage uncertainty is 0.02 / 0.13 = 15%, the same as 1 / 6.5 for the stack, but the absolute uncertainty is far smaller than the ruler resolution.Five timings of a trolley are 2.31 s, 2.40 s, 2.28 s, 2.35 s and 2.36 s. Find the mean time and an estimate of the uncertainty from the spread.
Worked examplesSum = 2.31 + 2.40 + 2.28 + 2.35 + 2.36 = 11.70 s, so mean = 2.34 s.Range = 2.40 − 2.28 = 0.12 s, so uncertainty ≈ half range = 0.06 s.Answer: (2.34 ± 0.06) s, a percentage uncertainty of 0.06 / 2.34 = 2.6%.
Section 7 of 8Key terms
| Term | Meaning |
|---|---|
| Resolution | Smallest change in a quantity that an instrument can display. |
| Range | Interval between the lowest and highest values an instrument can measure. |
| Accuracy | Closeness of a measurement to the true value. |
| Precision | Closeness of repeated measurements to each other. |
| Zero error | A systematic error where the instrument does not read zero for a true zero. |
| Parallax error | Apparent shift in a reading caused by viewing the scale from an angle. |
| Percentage uncertainty | Absolute uncertainty divided by the measured value, times 100%. |