Medical physics and astronomy (A Level)
24. Medical Physics
5 min read 10 parts Lesson 21 of 22 Unit quiz
Lesson summary
24. Medical Physics — Cambridge AS & A Level Physics 9702 — CAMBRIDGE AS & A LEVEL PHYSICS 9702 — COMPLETE NOTES notes: Explain how ultrasound, X-rays, CT scans and PET scans are produced, used and compared, and use acoustic impedance, attenuation and annihilation.Transducer = an…
- Piezoelectric effect
- Applications
- Transducer structure
- Transmit and receive
- Imaging process
- Acoustic impedance Z
Section 1 of 10The piezoelectric effect and transducers
- Transducer = any device that converts energy from one form to another.
- Piezoelectric effect = certain crystals (such as quartz, a lattice of silicon dioxide) produce a p.d. when deformed,
- and deform when a p.d. is applied.
- Compressing or stretching the crystal induces an e.m.f., and alternating compression gives an alternating e.m.f.
- An alternating p.d. makes the crystal vibrate at the same frequency, producing sound.
- The crystal is cut to a size that resonates at the required frequency.
- Applications
- microphones and ultrasound transmitters and receivers.
- (pressure variations to electrical signals)
- microphones and ultrasound transmitters and receivers.
Section 2 of 10Generating and using ultrasound
- Ultrasound = sound with a frequency above the range of human hearing (above 20 kHz), typically in the MHz range for medical use.
- Transducer structure
- a piezoelectric crystal with electrodes supplying an alternating p.d.
- Damping
- the crystal is backed with epoxy resin so it gives short pulses, which improves resolution.
- Transmit and receive
- the same crystal acts as transmitter (p.d. to sound) and receiver.
- (sound to alternating p.d.)
- the same crystal acts as transmitter (p.d. to sound) and receiver.
- Imaging process
- the transducer sends a pulse into the body, which is partly reflected at boundaries between tissues.
- The echo returns and generates an electrical signal for the scanner.
- The scanner uses the speed of sound and the time delay to calculate depth, and the echo intensity to show the nature of the boundary.
- Many pulses are combined into a two-dimensional image, as for an unborn child.
- Frequency
- a higher frequency gives a shorter wavelength and better resolution, so smaller details are seen.
Section 3 of 10Acoustic impedance and coupling gel
- Acoustic impedance Z = the product of the density of a medium and the speed of ultrasound in it, Z = ρc, in kg m−2 s−1.
- Sound is fastest in solids and slowest in gases, because closer particles pass vibrations faster.
- The c is the speed of sound, not light.
- Boundary behaviour
- part of the wave is reflected and part transmitted, and a larger difference in impedance gives greater reflection.
- Intensity reflection coefficient
- α = IR/I0 = (Z2 − Z1)2/(Z2 + Z1)2, given in the data booklet.
- If Z1 = Z2 there is no reflection.
- Energy is conserved, I0 = IT + IR.
- Problem
- air has Z about 400 kg m−2 s−1 and skin about 1.7 × 106, so almost all ultrasound reflects at an air gap (α ≈ 1).
- Coupling gel
- a gel with impedance similar to skin fills the gap, so very little is reflected (α ≈ 0.001) and most of the beam enters the body.
- It matches impedances and does not make the ultrasound travel faster.
Section 4 of 10Attenuation of ultrasound
- Attenuation = the loss of intensity as a wave is absorbed while travelling through a material.
- Equation
- I = I0e−μx, where μ is the absorption coefficient (m−1) and x the distance; this equation is not given.
- Frequency dependence
- about 0.5 dB cm−1 is lost for every 1 MHz of frequency.
- Image formation
- attenuation is not a major problem, because the image is made from reflections at boundaries.
- Echoes from depth
- the beam is attenuated going in, partly reflected and attenuated coming back, so the path is 2x.
Section 5 of 10Producing X-rays
- X-rays = short-wavelength electromagnetic radiation, about 10−8 to 10−13 m.
- Production
- electrons from a heated cathode are accelerated to a metal anode target.
- (thermionic emission)
- The fast electrons decelerate in the target, transferring kinetic energy to X-ray photons.
- Electrons in the target atoms dropping to lower energy levels release characteristic X-ray photons.
- electrons from a heated cathode are accelerated to a metal anode target.
- Maximum photon energy
- Emax = eV = hfmax = hc/λmin.
- Spectrum
- continuous,
- because electrons decelerate by different amounts, with a sharp cut-off at short wavelength because each electron gives its energy to at most one photon.
Section 6 of 10Quality of X-ray images and attenuation
- Aim
- minimise the patient's exposure and maximise the contrast and sharpness of the image.
- Exposure
- X-rays are ionising and can damage tissue or cause cancer.
- An aluminium filter absorbs long-wavelength (soft) X-rays, which are absorbed by the body and do not form the image.
- Contrast = the difference in blackening between structures.
- It improves with hard X-rays for bone, softer X-rays for tissue and a contrast medium.
- Good contrast needs a large difference between the transmitted intensities through different structures.
- Sharpness = how well-defined edges are, improved by a narrower beam, a collimator or lead grid to cut scattering, and smaller pixels.
- Attenuation
- bone absorbs X-rays strongly, so it appears white.
- I = I0e−μx, where μ is the linear absorption coefficient and depends on photon energy and material.
- Half-value thickness is the thickness of material that halves the beam intensity.
- For several materials multiply the factors, I/I0 = e−μ1x1 × e−μ2x2.
Section 7 of 10CT scanning
- CT (CAT) scan = an X-ray tube rotates around a stationary patient, taking images of one slice from many angles,
- and a computer combines successive slices into a 3D image that can be viewed from any angle.
| Advantages | Disadvantages |
|---|---|
| Higher resolution than a single X-ray | Much higher radiation dose |
| Distinguishes tissues with similar attenuation coefficients | Possible side effects from contrast media |
| Produces a 3D image |
- Dose
- a CT scan needs many exposures from many angles, whereas a plain X-ray needs one.
Section 8 of 10Radioactive tracers
- Tracer = a substance containing radioactive nuclei, introduced to the body and absorbed by tissue to study the structure and function of organs.
- Suitable isotopes
- technetium-99m and fluorine-18 bind to organic molecules, emit gamma or positron radiation and decay to stable isotopes.
- (such as glucose)
- Technetium-99m has a half-life of 6 hours.
- Fluorine-18 has a half-life of 110 minutes, so exposure is short.
- technetium-99m and fluorine-18 bind to organic molecules, emit gamma or positron radiation and decay to stable isotopes.
- Uses
- diagnosing cancer, finding areas of reduced blood flow in the heart, and detecting brain injury, Alzheimer's disease and dementia.
- Monitoring
- the tracer appears on the screen once tissues absorb it, showing the progress of a disease and the effect of treatment.
Section 9 of 10PET scanning and annihilation
- PET (positron emission tomography) = imaging of tissue by measuring the metabolic activity of cells, commonly using fluorodeoxyglucose (glucose containing fluorine-18).
- Decay
- fluorine-18 undergoes β+ decay: 189F → 188O + 0+1e + νe.
- Annihilation
- a positron travels less than a millimetre, meets an electron and both are destroyed, converting their mass to energy as two gamma photons in opposite directions.
- Mass, energy and momentum are conserved, and any particle and its antiparticle can annihilate.
- Each photon has E = hf = mec2 and momentum p = E/c.
Section 10 of 10Detecting gamma rays in PET
- Scanner
- the patient lies in a ring of detectors, each with a scintillator crystal and a photomultiplier.
- A gamma photon excites electrons in the crystal, which excite others, and as they return to lower states they emit faint visible light.
- The photomultiplier amplifies the light and converts it to an electrical signal.
- Line of response
- the two photons travel in opposite directions, so a pair of simultaneous hits defines a line through the annihilation point.
- Image
- a computer processes many lines and arrival times, and more photons from a point mean more tracer there, which appears as a bright area.
- Practical points
- the short half-life gives low exposure but the patient must be scanned quickly, and PET scanners are expensive and not in every hospital.
Last updated: