Part 1 of 6The Four Stages
Ions travel from the sample to the detector, and the time taken depends on their mass.
1
Ionisation: the sample is turned into positive ions, by electron impact or by electrospray ionisation.
2
Acceleration: the positive ions are accelerated by an electric field so that all of them have the same kinetic energy.
3
Ion drift: the ions pass through a field-free flight tube, where lighter ions travel faster and arrive sooner.
4
Detection: the ions hit a detector that records the time of flight and produces a current signal.
5
Data processing: a computer converts the flight times to m/z values and the signals to relative abundances in a mass spectrum.
Part 2 of 6Ionisation Methods
Electron impact
- Electron impact High-energy electrons knock an electron from each atom or molecule: X(g) + e⁻ → X⁺(g) + 2e⁻.
Electrospray
- Electrospray The sample is dissolved in a volatile solvent and sprayed through a charged needle, forming ions by gaining a proton.
Electrospray equation
- Electrospray equation X(g) + H⁺ → XH⁺(g), so the measured m/z is one unit greater than the relative molecular mass.
Vacuum
- Vacuum The instrument is kept under vacuum so ions do not collide with air particles on their way to the detector.
Part 3 of 6Why Mass Affects Flight Time
Why Mass Affects Flight Time
Equal kinetic energy
- Equal kinetic energy After acceleration, ions have the same kinetic energy, KE = ½mv².
Speed
- Speed Rearranging gives v = √(2KE ÷ m), so lighter ions have a higher speed.
Flight time
- Flight time The time to cross a flight tube of length d is t = d ÷ v = d√(m ÷ 2KE), so t rises with mass.
m/z
- m/z The mass spectrum plots abundance against mass-to-charge ratio, and for ions with a 1+ charge m/z equals the mass number.
Part 4 of 6Using the Spectrum
Peak position
- Peak position Each peak’s position gives the m/z of an isotope; its height gives the relative abundance.
Molecular ion
- Molecular ion For a molecule, the peak with the highest m/z is usually the molecular ion, giving the relative molecular mass.
Electrospray correction
- Electrospray correction If the spectrum was recorded by electrospray, subtract 1 from the highest m/z to find Mr.
Calculation
- Calculation Ar is found by multiplying each m/z by its abundance, adding, and dividing by the total abundance.
Part 5 of 6Worked Example
Question
- A sample of magnesium gave peaks at m/z 24, 25 and 26 with abundances 78.6 %, 10.1 % and 11.3 %.
- Calculate the relative atomic mass of magnesium to 3 significant figures.
- Multiply
- 24 × 78.6 = 1886.4;
- 25 × 10.1 = 252.5;
- 26 × 11.3 = 293.8.
- 24 × 78.6 = 1886.4;
- Add1886.4 + 252.5 + 293.8 = 2432.7.
- Divide by the total abundance of 100: 2432.7 ÷ 100 = 24.327.
- Round to 3 significant figures.
Answer Ar of magnesium = 24.3
Part 6 of 6Exam Checklist & Tips
1.1.3 – Key Points for the Exam
- m/z
- The mass-to-charge ratio of an ion;
- for a 1+ ion it equals the ion’s relative mass.
- The mass-to-charge ratio of an ion;
- Time of flight The time an ion takes to travel the length of the flight tube.
- Electrospray ionisation A soft ionisation method in which the sample is sprayed in solution and ionised by gaining H⁺.
- Molecular ion The ion formed when a molecule loses one electron, whose m/z equals the relative molecular mass.
Common Mistakes
- Saying heavier ions travel faster; at equal kinetic energy lighter ions are faster and arrive first.
- Forgetting to subtract 1 from the m/z of the molecular peak when electrospray ionisation has been used.
- Omitting the vacuum and stating that ions would be deflected by air in the tube.
- Treating the height of the peak as the m/z value rather than the abundance.